A derived characteristic value in test bench construction is the heat output from mass flow, specific heat capacity and temperature difference:
For a pure product, the relative uncertainties add in quadrature. The sensitivity coefficients cancel out[2].
The curve shows: the uncertainty of the temperature difference is an absolute quantity, and the spread is in the denominator. At a spread of 5 K and an uncertainty of 0.1 K, this gives 2 % on the output. This is more than mass flow and fluid property together. At 30 K, it is 0.33 %.
Correlation of sensor pairs
Flow and return sensors with a common calibration against the same standard and through the same input card have correlated deviations. The covariance term of the propagation law then decreases the uncertainty of the difference by the factor √2, compared with the calculation without correlation[2]. For this reason, legal metrology certifies sensor pairs and not single sensors.
The technical guideline for the verification of heat meters shows the system behind this[37]. It divides the meter into the 3 terms of the model: flow sensor, calculator and temperature sensor pair. For each term it gives a maximum permissible error and adds these errors arithmetically. Maximum permissible errors are not uncertainties. The 2 terms that depend on the temperature contain the ratio of the smallest spread to the measured spread. Thus they show the dominance from Fig. 2[37].
The same guideline documents a rule that plant construction usually gives on only by word of mouth. The measurement uncertainty of the test equipment must not be more than a specified fraction of the maximum permissible error at the test points. In accordance with the design, this is one fifth or one third[37].
The document EA-4/02 of European Accreditation (EA) contains fully calculated budgets as examples and is freely available[13]. The examples include the calibration of a thermocouple at 1 000 °C, of a temperature block calibrator at 180 °C and of a water meter.
For calorimetric measurements, there is a freely accessible report of the National Institute of Standards and Technology (NIST). It covers a 20 MW calorimeter and fully calculates 2 models with tabulated contributions in per cent[38].